hi and welcome to this training video on
troubleshooting the ultra care xt bed
frame the objectives for this lesson are
to demonstrate ways for troubleshooting
the ultra care xt bed frame in order to
isolate the failure and ensure the
correct fix is made within this lesson
we will troubleshoot the control box
motors and staff control in addition we
will look at proper bracket installation
and other helpful tips that will make
your ultra care xt bed perform exactly
as it was designed to do what do you do
when a bed fails to articulate at one
single point the ultra care xt has four
motors that are all controlled within
the control box which is where our
troubleshooting will begin and end if a
motor fails to operate then the problem
either stems from a bad motor or faulty
wiring within the control box so to
isolate the problem let's first trace
the line running from the motor to the
control box or simply look in the users
manual which lists lines one through
four the control harness is green and
additional plugs that go into the
control box include the emergency
battery pack socket and the AC power we
will first test to see if the problem
stems from one of the motors or if the
problem is within the control box begin
by unplugging the bed and then remove
the cover on the control box this can
easily be done with a screwdriver next
let's test the circuit to see if it is
working to do this unplug one of the
motors that is working correctly and
place it in the outlet that is failing
to operate this process will help to
diagnose whether the motor is bad or
if the circuit within the control box is
what's causing the problem if the motor
fails when it is plugged into another
outlet well the problem is with the
motor but if it works then the motor is
good and the trouble stems from within
the control box
you can test any motor in either of the
two high-low ports but the high-low
motors they can only be tested properly
in the high-low ports to replace the
control box we will need to disconnect
all four motors the control harness the
battery cable and the AC power cord to
disconnect the battery cable
use a screwdriver to push down on the
red tab and pull out on the cord to
release the tabs to remove the auxiliary
battery cable simply pop the C clamp up
using a screwdriver and unplug the cable
then remove the c-clip from the motor
and slide the control box off of the
motor now when it's time to reinstall
the control box simply slide it back on
the frame and secure it with the c-clip
begin by plugging in the battery cable
and secure it with the smaller c-clip
plug in the AC power cord then all four
motors in the control box if you need to
replace one of the high/low motors let's
first raise the bed to its highest
position if possible and unplug the
power cord from the wall outlet remove
the head and foot panels so that the
frame can be tipped on its side using
the needlenose pliers remove the rulings
and slide the clevis pins out of each
end of the actuator taking note of the
orientation of the pins as they must be
reinstalled in the same direction
discard used rue rings and clevis pins
you will then need to disconnect the
cable running to the control box when
checking the staff control first ensure
that the staff control panel has not
been damaged if
nothing is visibly damaged than the
electronics running from the control box
they are going to need to be checked to
do this disconnect the wires running to
the staff control and simply swap that
panel with another panel in your
facility if the swamp staff control
works well then the problem is with the
circuit board which is going to need to
be replaced however if you swap the foot
panels and nothing works then you will
need to check the patch cord which runs
to the staff control and if that checks
out fine then the only thing left is the
control box therefore that is what is
going to need to be replaced the foot
and head brackets they can be placed
incorrectly therefore let's review the
correct steps for placing both of these
brackets first lay the panel on a soft
surface the head panel bracket is the
longer of the two sizes then align the
top two holes with the bracket the
bracket offset should face up use the
four screws to install the brackets do
not tighten until after fitting the
mounting brackets into the bed sockets
repeat the same process with the foot
bracket panels panels should slip easily
in and out of their sockets another
helpful tip to include might be
lubricating rivet points periodically or
if your point squeak well thank you all
for your time and if you have any
additional
[Music]
Jorns a leading provider of solutions
for advanced wound management safety and
prevention and healthcare environment
design we collaborate and partner with
thought leaders and medical
professionals to bring a diverse
portfolio of services and products that
improve outcomes we are dedicated to
helping health care professionals in all
care settings enhance patient care
increase efficiency and deliver results
[Music]